Construction and description
In this range, temperature and humidity can be combined as you please, without resulting in
condensation. If there are extreme
ambient conditions, the working range may
If the specified range is exceeded upwards, e.g. 80 % rh at 60°C, the hot steam fed in will im-
mediately condense, due to the
dew-point, at the coldest point in the appliance.
At low temperatures and low relative air humidity, the effective range is heavily dependent on
the degree of humidity of the chamber load.
For the outer housing, MEMMERT deploys stainless steel (Mat.No. 1.4016 – ASTM 430) and
for the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through
its high stability, optimal hygienic properties and corrosion-resistance towards many (but not
all!) chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with
the materials mentioned. A material resistance table can be requested from the manufacturer.
Operating voltage and current consumption: See nameplate
Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010
Protection type IP 20 acc. to EN 60 529
Interference suppression acc. to EN 55011 class B
Appliance fuse: Fusible link 250 V/15 A quick-blow
The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V)
This appliance is intended for operation on an electrical power system with a system imped-
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
max
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
max
must ensure that the appliance is operated only on an electrical power system that meets
these requirements. If necessary, you can ask your local energy supply company what the
Observe the country-specific regulations when connecting (e.g. in Germany DIN VDE 0100
with residual current circuit breaker).